Yes, for the purpose of supply in Australia. Since 26 September 2022, Australian register entry 396331 has listed PARO’s maker, Intelligent System Co Ltd of Japan, as manufacturer of a Class I medical device described as a psychophysiological biofeedback system. Alpha Lifecare Pty Ltd is the sponsor. Because the entry is for a kind of device, PARO’s name is not on it, and a buyer should get written confirmation from the supplier that this entry applies to their unit. In the United States, PARO is listed with the FDA as a Class II biofeedback device that is exempt from premarket submission, which is different from FDA approval or clearance.
Procurement officers, biomedical engineers, funders and ethics committees ask this question for different reasons. This page sets out what the public records say, what they do not say, and how to check them yourself.
What is on the ARTG entry that covers PARO?
The Therapeutic Goods Administration publishes a one-page public summary for each entry on the Australian Register of Therapeutic Goods (ARTG). The summary for entry 396331, which we downloaded in October 2026, reads as follows.
| Field | What the public summary says |
|---|---|
| ARTG ID | 396331 |
| Entry name | Alpha Lifecare Pty Ltd - Psychophysiological biofeedback system |
| Entry type | Medical Device Included Class 1 |
| Sponsor | Alpha Lifecare Pty Ltd, postal address in Silverwater NSW |
| Manufacturer | Intelligent System Co Ltd, Nanto, Toyama, Japan |
| Start date | 26 September 2022 |
| Status | Active |
| Product type | Single Device Product |
| GMDN term | 63697 Psychophysiological biofeedback system |
| Specific conditions | None recorded |
The brand name is missing for a reason. The TGA explains that for lower risk classes, a “kind of medical device” covers similar products that share a sponsor, manufacturer, classification and Global Medical Device Nomenclature (GMDN) code, and that one entry may cover several models or brands. One cross-check supports the link to PARO: the US device identifier record for PARO Robot model MCR-900 carries the same GMDN code, 63697.
What does the intended purpose say?
The public summary gives the intended purpose in these words:
Intended to improve certain medical conditions through user interaction, including anxiety, depression, pain, loneliness, mood and/or motivation; and for symptoms such as agitation, aggression, walking about found in some mental health disorders and/or neurocognitive disorders such as dementia.
Read this as the purpose the device is supplied for. The TGA classifies medical devices by their intended purpose, and its overview of medical device regulation says the manufacturer is responsible for classifying its own device. The wording is not a summary of trial results. For those, the Queensland trial page sets out what the largest Australian study found and did not find.
What does Class I inclusion mean, and what does it not mean?
Under the Therapeutic Goods Act 1989, a kind of medical device must be included in the ARTG before it is supplied in Australia unless an exemption applies. The TGA’s overview makes these points about the class and the process:
- Class I is the lowest of the risk classes, with examples such as surgical retractors and tongue depressors.
- The sponsor is whoever carries legal responsibility for supplying the device here.
- The checking a device goes through rises with its class. For low risk devices it can be the manufacturer’s own self-assessment.
- For a Class I device that is not sterile and has no measuring function, the sponsor sends the manufacturer’s Declaration of Conformity with the application. Other documents, such as labelling, instructions and evidence of how the device performs, must be available to the TGA on request.
So an ARTG entry is the legal basis for supply. Before supplying, the sponsor must be able to show the device meets the TGA’s Essential Principles, so that it is safe and performs as intended. The entry does not show that the TGA has weighed the clinical evidence for reduced agitation or improved mood, and it is not an endorsement.
What is PARO’s status with the FDA?
Three US public records describe PARO, and they agree.
| Record | What it shows |
|---|---|
| FDA registration and listing database | “PARO Robots” listed by Intelligent Systems Co Ltd of Japan and by PARO Robots U.S. Inc., product code HCC, “Device, Biofeedback”, device class 2, under 21 CFR 882.5050 |
| AccessGUDID device record | PARO Robot, model MCR-900 UC Pink, product code HCC, marked exempt from premarket submission, with no premarket submission number |
| PARO Robots U.S. press release, 4 November 2009 | Describes receipt of a “Class II exempt pre-marketing notification” from the FDA on 21 September 2009 |
Put together, PARO is a listed Class II device that its US device record marks as exempt from premarket submission. Neither database shows a premarket approval or clearance number, so “FDA approved” and “FDA cleared” both describe it inaccurately.
How can you check the ARTG yourself?
The steps:
- Open the TGA’s ARTG search and search for 396331, or for the sponsor’s name.
- Open the entry and download the public summary PDF.
- Check the manufacturer, GMDN term, intended purpose and status. Status should read Active.
- Note the date printed on the summary. Its footer says it is not an ARTG certificate and that the reader should confirm the details are still current after that date.
- Get the supplier to name, in writing, the ARTG entry for your unit. File that number with the model and serial number.
If a supplier quotes a different entry number, search that number too. Because a kind of device is grouped by sponsor, another sponsor would need its own entry.
What this means for procurement and for claims
These records matter most when you write a business case, a grant application or a newsletter. A short guide:
| Wording | Use it? |
|---|---|
| “Included on the ARTG as a Class I medical device, entry 396331, sponsor Alpha Lifecare Pty Ltd” | Yes, once the supplier confirms the entry covers your unit |
| “TGA approved” | Use “included on the ARTG” instead, which is the register’s own term |
| “FDA approved” or “FDA cleared” | Avoid. PARO is listed and exempt from premarket submission |
| “Proven to reduce agitation” | Avoid. Quote the trial result, including the scale that showed no difference |
| “A medical device, so biomedical engineering must accept it” | Your hospital’s equipment program decides how it is classed and checked |
For a public hospital, the procurement checklist covers acceptance, electrical testing and the paperwork pack. For the company behind the device and how units reach Australia, see who makes PARO. Price, warranty and trial questions for a distributor are on the PARO buying page.
Next step
Ask for a PARO quote that names the ARTG entry for the unit supplied, so your records match the device that arrives.
Frequently asked questions
- Is PARO approved by the TGA?
The TGA's word for this is included. A Class I psychophysiological biofeedback system made by PARO's maker has been included on the Australian Register of Therapeutic Goods since 2022 under entry 396331. Inclusion is what allows it to be supplied here, and it is not a TGA finding about how well PARO works.
- Does the ARTG entry name PARO?
No. Lower risk entries are recorded by kind of device, grouped by sponsor, manufacturer, class and nomenclature code, so a brand name is not needed. Ask the supplier to state in writing that entry 396331, or another entry it names, covers the unit it is selling you.
- Is PARO FDA approved or FDA cleared?
Neither. US federal databases show PARO listed under product code HCC as a Class II biofeedback device and marked exempt from premarket submission. There is no premarket approval or clearance number on either record.
- Does ARTG inclusion prove that PARO works?
No. For a low risk Class I device the manufacturer can assess conformity itself, and the sponsor lodges the manufacturer's declaration of conformity with its application. The entry permits supply. Evidence about effects comes from trials, and the maker's own manual notes that results vary between users and are not guaranteed.
- Why does the maker's manual call PARO an electric toy?
The cover of the US user manual carries a caution label for electric toys and warns that care is needed to prevent fire or personal injury. That label sits on a US document. In Australia the ARTG entry is the record that matters, so ask the supplier for the manual issued with Australian units.
- Does a robotic companion cat need to be on the ARTG?
That depends on what its maker says it is for. The TGA's summary describes medical devices as products intended to diagnose, prevent, monitor, treat or alleviate a disease or injury, among other purposes. We have not checked companion pet listings, so search the ARTG by the maker's name before relying on any health claim in its advertising.