Public hospital procurement checklist for a therapy robot

A 14 step checklist for buying PARO through a public hospital: ARTG check, written quotes, AS/NZS 3551 acceptance, electrical testing and handover.

Public hospital procurement checklist for a therapy robot

A public hospital buys PARO the way it buys other ward equipment: a clinical case, product checks, a written quote, procurement, acceptance by biomedical engineering, electrical testing, infection control sign-off and a handover to the ward. The 14 steps below put that in order for a nurse unit manager, clinical lead or biomedical engineer. Your health service’s procurement policy and its medical equipment management program set the detail, so treat this as a list of what to have ready, not a replacement for either.

Before procurement: what is the clinical case?

1. Write the clinical case. Name the patients the robot is for, such as older patients living with dementia on a geriatric or aged care ward, or patients in a palliative care unit. Say who will run sessions, how often, and what you will record. The hospital and palliative care page describes how wards use therapy robots, and the acute wards page covers delirium and short stays.

2. Name the owner. Decide which ward or cost centre holds the asset, who looks after it day to day and who approves repairs. If a hospital foundation is paying, agree in writing that the unit becomes hospital equipment on arrival. The foundation funding page lists what a funder usually wants to see.

3. Book infection prevention and control early. PARO cannot be washed, according to the maker’s manual, so your infection control lead needs to approve a cleaning method for a shared soft-furred item before the first session. Send them the maker’s cleaning instructions as soon as you have them. The infection control checklist lists the questions they will ask.

Which supplier and product checks come first?

4. Check the ARTG entry. The TGA says medical devices must be included in the ARTG before they can be lawfully sold in Australia. Entry 396331 is a Class 1 inclusion for a psychophysiological biofeedback system. Its manufacturer is Intelligent System Co Ltd, PARO’s maker, and its sponsor is Alpha Lifecare Pty Ltd. The entry does not use the name PARO, so ask the supplier to state in writing which ARTG entry covers the unit you are buying, and record that number. The medical device page explains the entry in more detail.

5. Ask for the Australian charger and manual. The manual published in the US lists a charger input of 120 V at 60 Hz, which is not Australian mains power. It says to use only the adaptor supplied in PARO’s box and warns that using the adaptor in another country can cause overheating and damage. Ask for the adaptor and user manual supplied for Australia. Mains equipment designed or marketed for household or personal use falls under the Electrical Equipment Safety System, and in-scope Level 2 and Level 3 items must carry the Regulatory Compliance Mark. Ask the supplier to confirm the adaptor’s status under that scheme and check the label when it arrives.

6. Get warranty and service terms in writing. No Australian warranty length or repair turnaround for PARO is published. Ask for the warranty, the repair route, loan unit terms, battery and fur replacement prices and how long parts will be supplied. The servicing and repairs guide lists 10 terms worth settling.

Procurement steps

7. Get a written quote. Ask for the unit price, GST, freight to the hospital, the warranty and the delivery lead time on one document. The directory figure for PARO starts at $6,500, but the distributor’s quote is the number procurement will use.

8. Follow your purchasing route. Your procurement team decides how many quotes are needed, whether the supplier must be registered as a vendor and how the purchase order is raised. A foundation purchase or donation still needs to reach the ward through that route, or through your gifts policy, so the unit is recorded properly.

9. Plan the trial question. If the ward wants to try PARO before committing, settle whether the trial unit is on loan from the supplier and who is liable for damage. AS/NZS 3551 covers equipment on loan or on trial as well, so a trial unit goes through acceptance too (step 10).

What does biomedical engineering need for acceptance?

10. Run acceptance under your equipment management program. AS/NZS 3551:2012, Management programs for medical equipment, sets out procedures for procurement, acceptance, fault management, routine testing and disposal. Standards Australia describes it as covering all medical equipment, whether hospital property, privately owned, on loan or on trial. The 2012 edition has two amendments, the second from 2016, and the store marks it as pending revision. Your program decides how PARO is classed and checked.

11. Hand biomedical engineering a complete pack. Include:

  • Model, serial number and ARTG entry number.
  • The user manual and the maker’s cleaning instructions.
  • Charger details and the battery specification. The US manual lists a 7.4 V lithium-ion battery pack.
  • Supplier contacts, warranty dates and the agreed fault route.

The serial number is printed on the sides of the box and under a lid on PARO’s underside. Each unit has its own, and the maker asks for it with any support request.

Electrical safety testing

12. Agree the test regime. PARO runs on its internal battery or from its mains adaptor, so the mains-connected parts are the adaptor and its cord. Biomedical engineering decides whether that is done under AS/NZS 3760:2022, the in-service safety inspection and testing standard for electrical equipment and RCDs published on 24 June 2022, or under the testing schedule in the hospital’s own AS/NZS 3551 program. Record the decision and the test interval in the asset register.

Handover to the ward

13. Train the ward before the first session. Cover these points from the manual:

  • The switch is between the hind flippers under the fur: press about 1 second to start, about 3 seconds to stop.
  • A full charge takes about 3 hours from empty, and PARO only charges while switched off.
  • Charge and store it on a flat, stable surface away from sunlight and heat, never on a bed or under a blanket.
  • For smoke, a strange smell, abnormal sounds, overheating or liquid inside: unplug the charger, switch PARO off, remove the battery pack and log the fault.

14. Review after the first months. Look at session logs, faults and any outcome measures, then decide whether to keep, extend or return the unit. Keep the packaging, because the maker says it is needed if PARO is returned.

Summary

Step Suggested owner Keep on file
1 to 3 Nurse unit manager or clinical lead Clinical case, owner, infection control referral
4 to 6 Nurse unit manager with procurement ARTG number, Australian charger confirmation, written terms
7 to 9 Procurement Quote, purchase order or gift record, loan agreement
10 to 12 Biomedical engineering Acceptance record, asset number, test regime
13 to 14 Ward Training record, session log, review notes

Request a hospital quote that includes the ARTG number and service terms.

Request a free quote

Frequently asked questions

Is PARO tested under AS/NZS 3551 or AS/NZS 3760?

Biomedical engineering decides. AS/NZS 3551 sets out how a hospital manages medical equipment, including acceptance and routine testing, and lists AS/NZS 3760 among its cited references. AS/NZS 3760 covers in-service inspection and testing of electrical equipment and RCDs. Ask biomedical engineering which regime it applies to PARO's mains adaptor, and record the regime and test interval in the asset register.

Does PARO have to be on the ARTG for a hospital to buy it?

The TGA says medical devices must be included in the ARTG before they can be lawfully sold in Australia. ARTG entry 396331 covers a Class 1 psychophysiological biofeedback system made by PARO's maker, with Alpha Lifecare Pty Ltd as sponsor, but it does not name PARO. Ask the supplier to confirm in writing which entry covers the unit you are buying.

What should biomedical engineering receive with the unit?

The model and serial number, the ARTG entry number, the user manual supplied for Australia, the maker's cleaning instructions, the charger details, the warranty dates and the supplier's contact for faults. Biomedical engineering then records PARO in the asset register and sets the test regime.

Can a foundation buy PARO and donate it to the ward?

Yes, but agree first that the unit becomes hospital equipment on arrival. It should still go through your gifts or purchasing policy, acceptance by biomedical engineering and infection control review, with a written quote attached to the funding request.

Why ask for an Australian charger and manual?

The US edition of the maker's manual lists a charger rated for 120 V at 60 Hz and warns against using the adaptor in another country, because it can overheat and damage PARO. Australian mains power differs, so ask for the adaptor and manual supplied for Australia and check the adaptor label when it arrives.